Programming the blinking decimal point. Press
key to situate it in the desired
position. If no decimal point is desired then you have to situate it to the righter place
in the display. The selected position will be fixed for all the programming and run
mode.
ENTER
Validate data and access to next programming step..
ESC
Cancel programming and returns to run mode.
Programming input value in point 2, led INP2 on.
Key-in method Use
to switch between "0" (positive) and "-" (negative). Press
to advance to the next digit to the right which goes in flash. Press repeatedly
to increment the active digit until it takes desired value. Proceed in the same
manner with all the digits until desired value is completed on the display with sign.
Press
ENTER
to accept this value as INP2 and go next step.
Teach method: Press
TEACH
key to display the real input value.
ENTER
Validate data and access to next programming step.
ESC
Cancel programming and returns to run mode.
Programming of the display value for the first point, activated LED DSP2. By means of
the
and
procedure, program desired DSP2 value and press
ENTER
. The
limits of the span are -9999 and 9999 points..
To save changes and return to run mode press
ENTER
key or,
To access the programming of the scale linearisation points, press
ENTER
key during 3
seconds.
ESC
Cancel programming and returns to run mode.
20
[20.1] Decimal point
B
A
TARE
HOLD
LIMIT
MAX
MIN
DATA
DSP2
INP2
FLT
STORE
DSP1
INP1
2
1
4
3
TARE
RESET
LIMIT
MAX/MIN
ENTER
ESC
DATA
PROG
TEACH
PRG
RUN
[20.2] Input 2 value
B
A
TARE
HOLD
LIMIT
MAX
MIN
DATA
DSP2
INP2
FLT
STORE
DSP1
INP1
2
1
4
3
TARE
RESET
LIMIT
MAX/MIN
ENTER
ESC
DATA
PROG
TEACH
PRG
RUN
[20.3] Display 2 value
B
A
TARE
HOLD
LIMIT
MAX
MIN
DATA
DSP2
INP2
FLT
STORE
DSP1
INP1
2
1
4
3
TARE
RESET
LIMIT
MAX/MIN
ENTER
ESC
DATA
PROG
TEACH
PRG
RUN
Warning: VERY IMPORTANT: Scaling the
meter with a tare value different from zero
may cause false readings when exiting to
the run mode. Before trying to program
the scale, check the TARE LED and, if
activated proceed to clear the tare
memory (Fig. 26.2).
40
RS232
ModBus-RTU protocol compatible (see ModBus manual).
RS485
Send data to a burster printer. New programming menu
("timE") that prints data and time, page 32.
ModBus-RTU protocol compatible (see ModBus manual).
ANALOG OUTPUT
See connector functions, page 32.
SAMPLE & HOLD
New useful
function to show and to process very quick
phenomenons, holding on display and on the outputs the
values registered just at the moment of activating this
function until this be deactivated. See pag.33
4.1 ADITIONAL FUNTIONS
The new 9181 of ±9999 points increases and enhances the
functions and programming of the following output options:
SETPOINTS
Setpoint value programming from +9999 to -9999.
1.
New functions :
1.1.
Setpoint biestables "latch". This setpoints once
activated, remains in this status until a external
reset of this function (see RESET setpoints nº25,
on the logical functions table, page 32). It ’s use
allows to know the previous activation of relais in
installations where the instrument is not constantly
controled.
1.2.
Setpoint activation by: net value, gross value, peak
value or valley value.
1.3.
Indication of Relay activation with LED in front or
additionaly flashing the reading when reached the
setpoint value.
Quick access to the setpoints value programming.