7 | Cleaning and servicing
BÜCHI Labortechnik AG
72/116
Operation Manual Rotavapor® R-300
1
2
3
Fig. 48:
System Overview
view
1
Serial number
2
Duty hours
3
Firmware
ð
The mobile device shows the
System Overview
view.
u
Verify that the serial number displayed matches the serial number on the type
plate.
u
Tap the
[ Firmware ]
box.
1
2
3
Fig. 49:
Update Firmware
view
1
Notification bar
2
[ Update ]
button
3
List of the components for which the
firmware is to be updated.
ð
The mobile device shows the
Update Firmware
view.
ð
The
Update Firmware
view shows a list of all the firmware updates that are
to be carried out.
u
Make sure that the notification bar shows
[ Firmware update ]
.
ð
NOTICE! All of the components are updated. Selecting individual com-
ponents is not possible.
u
Tap
[ Update ]
.