BÜCHI Labortechnik AG
Installation | 5
Operation Manual Rotavapor® R-300
45/116
5.17.3
Overview: setting up coolant tubing connections
The tubing connections between the various items of BUCHI laboratory equipment
form a sealed circulation system. The starting and finishing point is always the
recirculating chiller (F-3xx).
Below is an example of the tubing connections between the laboratory apparatus.
Recirculating
Chiller F-3XX
Rotavapor
R-300
Vacuum Pump
V-300 / V-600
2
3
6
4
1
5
Fig. 26: Coolant tubing connections in a BUCHI distillation system (example)
1
Inlet on Recirculating Chiller F-3xx
4
Condenser outlet on Rotavapor
R-300
2
Outlet on Recirculating Chiller F-3xx
5
secondary cold trap inlet on Vacuum
Pump V-300
3
Condenser inlet on Rotavapor R-300 6
Secondary cold trap outlet on Vac-
uum Pump V-300
u
Connect a tube between the outlet of the recirculating chiller (2) and the inlet of
the condenser on the Rotavapor R-300 (3).
u
Connect a tube between the outlet of the condenser on the Rotavapor R-300 (4)
and the inlet of the secondary cold trap on the Vacuum Pump V-300 (5).
u
Connect a tube between the outlet of the secondary cold trap on the Vacuum
Pump V-300 (6) and the inlet of the recirculating chiller (1).
NOTE
Use GL14 hose barbs for the tubing connections.
Secure tubes with spring clips where necessary.
NOTE
The R-300 is also suitable for use with water coolant.