6 | Operation
BÜCHI Labortechnik AG
52/116
Operation Manual Rotavapor® R-300
6.1.4
Fitting the receiving flask
1
2
Fig. 32: Fitting the receiving flask
1
Ball joint clamp
2
Receiving flask
u
Fit the receiving flask (2) onto the condenser from below, hold it firmly in place and
secure it with the ball joint clamp (1).
u
If the connection is not sufficiently airtight, apply a thin coat of Glisseal 40 red
laboratory grease to the ground glass joint.