BÜCHI Labortechnik AG
Operation | 6
Operation Manual Rotavapor® R-300
49/116
6
Operation
6.1
Preparations
6.1.1
Preparing the heating bath
CAUTION
Risk of skin burns from hot fluids and surfaces
u
Do not put your hands in the hot fluid.
u
Do not carry, shift, tip or otherwise move the heating bath when it is filled with hot
fluid.
u
Do not fill the heating bath right to the very top. Allow for the expansion of the fluid
when heated.
u
Do not switch on the heating bath when it is empty.
CAUTION
Risk of skin burns from oil splashes
u
Never pour water into hot oil.
u
Make sure that the oil used is suitable for these purposes and the desired temper-
atures. (The oil used must have a flashpoint of > 175 °C.)
on
Lift set
max.95°
46
°C
50
°C
Set
STOP
1
2
5
1
9
7
6
3
4
8
Fig. 29: Heating bath display
1
Maximum temperature
6
Set temperature locked
2
"Lift set" indicates that a lower limit
has been set for the Rotavapor lift.
7
Connection symbol – heating bath is
being controlled by an Interface I-300/
I-300 Pro
3
Actual temperature
8
START/STOP function activated on
pressing navigation control
4
Set temperature
9
Control knob for setting the required
temperature and starting/stopping the
heating process
5
ON/OFF master switch