Büchi Labortechnik AG
Operation | 6
Operation Manual Pure Chromatography Instruments
37/102
6.1.4
Conditions panel
1
1
Information button
(Information about the selected col-
umn)
The conditions area contains the following settings:
Action
Option
Explanation
[Column]
(Prep mode)
Enter value
Enter a name for the present car-
tridge.
[Column]
(Flash mode)
Select column /auto-
matic via RFID
Shows the name of the column in-
stalled in the instrument.
[Flow Rate]
Enter value
Edit the default flow rate.
[Duration Units]
Choice of the duration
type
The following types are available:
minutes / column volumes
[Equilibration]
Enter value
Indicates the period of time or num-
ber of column volumes that the mo-
bile phase flows through the column
before the sample is injected.
[Run Length]
Enter value
According to the current operation
enter the time for the separation.
According to the current operation
enter the number of column volumes
required for the separation.