6 | Operation
Büchi Labortechnik AG
62/102
Operation Manual Pure Chromatography Instruments
6.5.2
Selecting the flash mode (Pure C-850 only)
Navigation path
➔
Mode
u
Navigate to the
Mode
menu according to the navigation path.
u
Select
[Flash]
.
ð
The flash mode is selected.
6.5.3
Starting Flash separation using a method
Precondition:
R
The instrument is prepared. See Chapter 6.5.1 "Preparing the instrument",
page 61
R
The required solvents are connected and assigned. See Chapter 5.7 "Assigning
solvent to solvent lines", page 26
R
The sample is prepared.
R
The waste bottle is empty.
u
Calibrate the solvent level sensor. See Chapter 5.8 "Calibrating the solvent level
sensor", page 27
u
Calibrate the waste level sensor. See Chapter 5.12 "Calibrating the waste level
sensor", page 29
u
Open the protection shield.
u
Place the fraction collection trays inside the instrument. See Chapter 6.4.5
"Inserting the fraction collection tray", page 60
u
Close the protection shield.
u
Open an existing method. See Chapter 6.4.6 "Selecting an existing method",
page 61
u
Tap the button
[OK]
.
u
Tap the button
[Start]
on the
Run
panel.
u
According the requirements adjust the file name.
u
Tap the button
[OK]
.
ð
The display shows the dialog box
Sample Loading
.
u
According the requirements adjust the settings. See Chapter 6.4.7 "Selecting
values on the dialog box Sample Loading", page 61
u
Follow the instructions on the display.
ð
Installing the cartridge. See Chapter 6.4.3 "Installing and removing a cartridge",
page 57
ð
Introducing the sample into the system. See Chapter 6.4.2 "Injecting a sample
into the flash system", page 55
6.5.4
Changings during a separation
NOTE
Parameters that can be edited are highlighted in green