Büchi Labortechnik AG
Cleaning and servicing | 7
Operation Manual Pure Chromatography Instruments
73/102
7
Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this
section.
Any servicing and repair work which involves opening up the casing may only be
carried out by BUCHI service technicians.
u
Use only genuine BUCHI consumables and spare parts in order to ensure correct
operation of the device and preserve the warranty.
7.1
Regular maintenance work
Component
Action
Frequency
Pumps and valves
u
Purge the instrument with air. See
Chapter 7.8 "Removing solvent from an
used cartridge", page 78
u
Purge the instrument with purging
solvent. See Chapter 7.6 "Cleaning the
instrument", page 77
Daily
Tubing
u
Check for leaks. If a leak is observed,
resolve the issue before continuing.
Daily
Fittings
u
Inspect fittings; if solid material is
deposited on a fitting, clean and
tighten (replace) the fitting before
continuing.
Daily
Drain lines
u
Check all drain lines to ensure that
liquid can flow through them to the
waste container
Daily
Waste bottle
u
Empty the waste bottle
Daily
Filters
u
Check the filters in the solvent bottles
and clean if necessary.
Weekly
Fittings
u
Check and if necessary tighten the
fitting that secures the tubing from the
mobile phase reservoir manifold to the
inlet check valve housing
Weekly
Data
u
Perform a data backup
Weekly
Casing
u
Wipe down the casing with a damp
cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Weekly
Warning symbols
u
Check that the warning symbols on the
instrument are legible.
u
If they are dirty, clean them.
Weekly
Display
u
Wipe down the display with a damp
cloth.
Monthly
All manuals and user guides at all-guides.com