8 | Help with faults
Büchi Labortechnik AG
88/102
Operation Manual Pure Chromatography Instruments
Error message
Possible cause
Solution
Instrument Alarm:
Vapor
Vapor sensor settings are set
too sensitive
u
Change the vapor limits to a
lower sensitivity. See
Chapter 6.2.9 "Editing the
vapor sensor sensitivity",
page 52
Leaks
u
Check for any leaks in the
solvent flow paths.
Solvent vapors in work area
u
Solvent vapors in the work
area may set on the alarm.
Use the instrument in a
hood or well ventilated area
with no open solvents near
the system.
Vapor sensor no sig-
nal
Vapor sensor is not connected
or is defect
u
Check the cable to the vapor
sensor.
Instrument Alarm:
Watchdog
Timing/communication issue
in software
If in setup mode:
u
Restart the instrument.
u
If in run mode:
u
Press reset.
u
Restart after the run is
finish.
Solvent Pressure
Solvent pressure limit has
been exceeded
u
Remove column.
u
Replace with bypass.
u
Determine if column is not
the source of blockage.
u
If not column is the source
of blockage, then check all
fluidic lines.