Büchi Labortechnik AG
Appendix | 10
Operation Manual Pure Chromatography Instruments
97/102
10.3
Finding out separation conditions with the navigator
10.3.1 Open the navigator
Navigation path
➔
Gradient panel
u
Navigate to the
Gradient
panel according the navigation path.
u
Tap the button
[Navigator]
.
ð
The display shows the dialog
Navigator
.
10.3.2 TLC Silica
The TLC-Silica tool uses TLC separation data to provide a recommended gradient for
silica separations of 2 or 3 components.
Specification of the samples needed:
Two different solvent concentrations
Two separations
Silica TLC plates
Precondition:
R
The display shows the dialog
Navigator
. See Chapter 10.3.1 "Open the
navigator", page 97
u
Select the
TLC- Silica
tab.
u
Select the radio button for two or three components.
u
Select the weaker solvent at drop-down list A.
u
Select the stronger solvent at drop-down list B.
u
Enter the solvent B concentrations used in the TLC separations.