6 | Operation
Büchi Labortechnik AG
52/102
Operation Manual Pure Chromatography Instruments
6.2.9
Editing the vapor sensor sensitivity
The vapor sensor detects solvent concentration in the ambient air.
The following sensitivity limits are available:
Sensitivity
Meaning
High
Used for non-volatile solvents
Medium
Compromise between the low and high setting
Low
Used for volatile or semi-volatile solvents
Off
The vapor sensor is off
Navigation path
➔
Tools
➔
Vapor Sensor and Limits
u
Navigate to the
Vapor Sensors
dialog according to the navigation path.
ð
The display shows the dialog box
Vapor Sensors
.
u
Tap the drop-down list next
[Vapor alarm sensitivity]
.
u
Select the sensitivity value you wish to use.
u
Tap the button
[Close]
.
ð
The dialog box closes.
6.2.10 Saving a method
Navigation path
➔
File
➔
Save Method as