During the start-up of the Remote mode different wireless nets
can be found present in the working area.
Always use the “BTSFREEEMG 300” net ignoring the other nets.
From the Workstation launch the application that you want to use (for
example Myolab.
Wait until, in the application being used, the respective LED indicating
the state of the Wi-Fi connection between the receiving unit and the
workstation changes from red to flashing green.
Correct data visualization check
During the initial connection, check the correct operation of the probes in
use, and the correct display of data on the BTS FREEEMG 300.
To do this, activate all the EMG probes and, if the system is also provided
with footswitches, two FSW/EGN probes, enabled as FSW, to be tested
all previously charged, as indicated in § ““Active” Button”. When all the
probes are shown to be active on the display, select the Remote mode
by clicking on the “Remote” button. Select the defaul protocol and then
“Setup Probes” from the menu and wait for all the activated probes to be
acknowledged (the respective box must be green).
To verify the correct functioning of the EMG probes, position a pair of
electrodes (see § “Fitting the Patient”) and connect Probe 1. Next, select
“Gains” from the menu and check the progress of the EMG1 signal.
Repeat the procedure connecting the other probes one at a time.
Содержание FREEEMG 300
Страница 1: ...BTS FREEEMG version 3 1 0e user manual english ...
Страница 2: ...Document Number ERFNC 00782 01e Published October 2010 Copyright 2010 BTS SpA All Rights Reserved ...
Страница 6: ......
Страница 108: ......
Страница 115: ...113 BTS Bioengineering BTS FREEEMG appendix E declaration of conformity ...
Страница 119: ......