FREEEMG
acquisition protocol
using the appropriate box “Description”
- make sure that the BTS FREEEMG 300 is in the Lab Modality (to
access this modality from the main page, click on the “Remote”
button; if two BTS FREEEMG 300 units are being used, make
sure that both the unit have the Remove Mode activated)
- click on “Write by WiFi” to save the protocol on the PC and
transfer it on to the BTS FREEEMG 300.
If two receiving units are being used, it is necessary to wait for
the protocol to be transferred to both units. At the end of the
transfer, the protocol will be visible in both, but in each unit will
be displayed only the respective signals from the channels they are
connected to.
For example, if a protocol with 10 EMG channels has been
created ( 8 associated with the first unit, 2 with the second) and
the footswitches, the protocol on the first unit will register 8
EMG channels and the FSWs, while the protocol on the second
unit will register only 2 EMG channels.
Only the enable signals (“Enable” checked ) will be memorized
in the protocol.
If for example a 4 signals plus FS protocol has already been defined
to create a protocol including only the 4 EMG signals it is enough
to remove the checks on “Enable” from the FS fields.
If required it’s possibile to transfer the protocol through a cable:
- connect the PC and the RU through a USB-PDA cable
- click on “Write Protocol” to save on PC and transfer the protocol
on the BTS FREEEMG 300.
Содержание FREEEMG 300
Страница 1: ...BTS FREEEMG version 3 1 0e user manual english ...
Страница 2: ...Document Number ERFNC 00782 01e Published October 2010 Copyright 2010 BTS SpA All Rights Reserved ...
Страница 6: ......
Страница 108: ......
Страница 115: ...113 BTS Bioengineering BTS FREEEMG appendix E declaration of conformity ...
Страница 119: ......