5.3 - Repair
M2041/5en
SBT-2400 Service Instructions
39
©BTG 2006
5.3 Repair
5.3.1
Transmitter Replacement
•
Before any service is carried out, the transmitter must be carefully cleaned
from pulp and other impurities.
•
If the transmitter is taken to a workshop for detailed service, ensure that
the workbench is clean. Particles can penetrate the differential
transformer and degrade the performance.
5.3.1.1
Remove the Transmitter from Process Pipe
1. Loosen the cable connection and remove the cable.
2. Loosen the clamp.
3. Remove the transmitter from the pipe and clean it.
5.3.1.2
Change the O-ring Seal
.
5.3.1.3
Mount the Transmitter
.
Tools required:
Spanner
DANGER
CHECK PIPES!
Risk for corrosive,
toxic, or aggressive
liquids or gases
with high pressure or
high temperature.