4.2 - Operations with JCT-1200
M2041/5en
SBT-2400 Operating Instructions
25
©BTG 2006
4.2.3.2
Multi-point Calibration
Start
Is
sample list
full?
Yes
No
Is
sample to be
used in
calculation?
Yes
No
Remove a lab sample.
See chapter “Remove
Sample Values”.
Use sample in
calculation. See
chapter “Calculate New
Calibration Constant”.
Done
Take a lab sample. See
chapter “Take New
Sample Values”.
Store the lab sample.
See chapter “Enter Lab
Sample Values”.