5.4 - Troubleshooting
M2041/5en
SBT-2400 Service Instructions
51
©BTG 2006
4. SBT-2400 oper-
ates in a normal
manner, except the
HART interface.
4.1. Transmitter software is not
responding
Reboot the transmitter.
Disconnect the cable to the transmitter. Wait 10 seconds.
Reconnect the cable.
Is HART working now?
If not, go to next problem cause.
4.2. Faulty electronics unit
Verify the following:
- The measuring resistor must be 250 ohm minimum.
- The supply voltage is high enough.
- The terminal voltage is 11V minimum
Does HART operate at any signal current?
If it does, try changing the cable.
If that does not work, or HART does not operate at all,
change the electronics unit.
Symptom
Probable Cause
Action