5.3 - Repair
44
SBT-2400 Service Instructions
M2041/5en
©BTG 2006
5.3.4
Measuring Module Replacement
The measuring module is replaced when:
•
The blade has been exposed to severe external forces and the measuring
module settings are distorted and can not be adjusted correctly.
•
The differential transformer is damaged.
•
The troubleshooting of the electronic indicates an error in the measuring
module.
NOTE!
The measuring module must not be repaired or adjusted! The module is
delivered as a complete spare part. A faulty measuring module MUST be sent
to BTG for factory repair.
It is worth sending a faulty measuring module to BTG for repair rather than dis-
carding the unit, as the measuring module is the most expensive part of the
transmitter. Contact your local BTG dealer.
5.3.4.1
Remove the Measuring Module
1. Remove the blade. See section 5.3.2.1:
Remove the Blade and Deflector
.
2. Remove the housing cover.
3. Remove the circuit board. See section 5.3.3.1:
4. Remove the 3 bolts (1) at the bottom of the transmitter housing.
5. Carefully extract the measuring module (2).
6. Send the measuring module to BTG for factory repair.
Contact your local BTG dealer.
Tools required:
Snap ring pliers
2.5 mm Allen key/hexagon spanner
4 mm Allen key/hexagon spanner
(Small pliers)
(Wise)
1
2
Содержание Pulptec SBT-2400
Страница 1: ...USER MANUAL Pulptec BTG Pulp Paper Sensors AB SBT 2400 Static Blade Consistency Transmitter M2041 5en ...
Страница 12: ...1 4 Type Sign Explanation 6 SBT 2400 Product Introduction M2041 5en BTG 2006 ...
Страница 26: ...3 6 Mounting of Optional Accessories 20 SBT 2400 Installation Instructions M2041 5en BTG 2006 ...
Страница 58: ...5 4 Troubleshooting 52 SBT 2400 Service Instructions M2041 5en BTG 2006 ...