8.3.3
Surveying the SMART X2S for Radiation Leakage
The SMART X2S should be surveyed for radiation leakage and stray radiation as a closed-beam X-ray
diffraction system in accordance with the following documents:
•
Guide for submission of information on Analytical X-ray Equipment required pursuant 21
CFR 1002.10.
•
National Bureau of Standards Handbook III Radiation Safety for X-ray Diffraction and
Fluorescence Analysis Equipment.
WARNING
!
CAUTION
!
Radiation surveys should only be performed
by—or under the immediate supervision
of—qualified Bruker AXS personnel.
X-ray Emission
1.
Perform the X-ray emission survey with the X-ray tube to be supplied with the
instrument, operated at a power level of 50kV and 1 mA to produce the maximum X-ray
exposure. (Reduce the mA as necessary so as not to exceed the maximum rated power
of the tube.) The test shall be conducted with the supplied optics.
WARNING
!
CAUTION
!
Individuals performing the survey should wear
leaded safety glasses as a precaution. Dosage in
excess of 2.5 mR/hr may be hazardous to the
eyes.