Introduction to the SMART X2S
SMART X2S User Manual
2 - 4
DOC-M86-EXS127 V4
2.3.3
Environment Requirements
Table 2.3 — SMART X2S environment requirements
NOTE:
The SMART X2S is designed for indoor
use only.
2.3.4
Other Specifications
Table 2.4 — SMART X2S other specifications
Environment Requirements
Installation surface
Horizontal tabletop, countertop, or other stable, vibration-free surface rated for the
weight of the SMART X2S. For additional stability, the SMART X2S may be bolted
to the installation surface.
Heat dissipation
less than 300 W (1024 btu/hr)
Clearance behind
at least 3” (7.6 cm) for proper airflow and ventilation
Clearance above
at least 3’ (.91 m) for service access
Altitude
up to 6562’ (2000 m) above sea level
Ambient temperature range
59-77°F (15-25°C)
Ambient humidity range
20%-80%, non-condensing
Specification
Value
SMART X2S Electronic Components Overvoltage
Category
II
SMART X2S Pollution Degree
2
X-ray Tube
50 W microfocus tube, molybdenum anode, air-cooled
X-ray Optics
Toroidally-curved Si crystals for three-dimensionally-focused,
monochromatic X-rays with very high flux density
CCD Detector
Air-cooled, no spatial distortion
Sample Mount Type
Bruker Spine Pin
Low-temperature Device (optional)
Supports DTC (Desktop Cooler) low-temperature device
Data Collection Resolution
To IUCr standards