User Manual
5 Step-by-Step Guides
98
5.15
Line Scan
This section describes the steps of line scan analysis using the Line scan workspace.
Step
Example/hints
1
Select the
workspace.
2
an image.
Refer to section 5.3.
3
Highlight the line and drag and adjust the
endpoints to the desired position.
4
Set
Point count
of the line scan.
Alternatively select
Distance [µm]
between
measurement points.
5
Click at
the
button to
set
Scan time
.
Choose
Automatic
,
Manual
or
Measurement time [s]
6
a line scan.
The line scan is automatically terminated if
Automatic
or
Measurement time [s]
is
selected. When
Manual
is selected,
acquisition can be stopped manually by
clicking the
button. One click
terminates scan after finishing the last scan
(the button changes into
), a
second click terminates the measurement
immediately.
7
Use the
icon to identify elements.
Elements are automatically identified by
Auto
ID
. Add and delete elements by clicking on
the
Element ID
or use the
Finder
in the
Spectrum tab. Identified elements are
displayed in the line profile thumbnail bar.
The element selection can be changed at any
time during or after the line scan acquisition.
8
Select elements in the thumbnail bar by
ticking the boxes below the individual
element images to display their profiles
in the
Profiles
tab.
Optionally use
or
on the right of
the thumbnail bar. Use the
icon to edit
the line profile display settings, e.g. scaling of
the x-axis, selection of result types, filter
strength, etc. (see also Fig. 5.16-1).
9
Use right mouse key in the scan image
to extract region of interest spectrum
from the line scan.
Extracted spectrum (named as
Range
)
appears in the
Spectrum
tab. (optional step).
The
Scan
spectrum in the spectrum tab
presents the sum spectrum of the line.
10
Use the workspace
icon to
a)
Save
line scan
Use .rtl file format to save line scan data
including EM and scan images and point
spectra.
b)
Add
data
to project
Image(s), composite element profiles and
scan image will be transferred.
Confirm the pop-up window
Do you want to
save point spectra too?
by clicking
to transfer spectral data.
c)
Add
data
to report.
Содержание QUANTAX EDS
Страница 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Страница 6: ......
Страница 8: ...User Manual 2 Safety Information 8 ...
Страница 12: ......
Страница 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Страница 22: ......
Страница 24: ......
Страница 48: ......
Страница 50: ...User Manual 5 Step by Step Guides 50 ...
Страница 52: ...User Manual 5 Step by Step Guides 52 Fig 5 1 1 Preparatory steps for measurements with the ESPRIT software ...
Страница 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Страница 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Страница 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Страница 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Страница 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Страница 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Страница 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Страница 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Страница 85: ...5 12 Quantitative EDS Analysis of Bulk Samples QUANTAX EDS 85 Fig 5 12 2 Automatic standardless quantification ...
Страница 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Страница 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Страница 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Страница 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Страница 116: ......