5.2 Energy-Channel Calibration
QUANTAX EDS
53
5.2
Energy-Channel Calibration
Correct energy-channel calibration is a prerequisite for reliable qualitative and quantitative analysis
results. Calibration should be checked on a regular basis or after re-start of the signal processing
unit (SVE).
Step
Examples/hints
1
Select the
workspace and
the
Spectrometer
tab.
Do not confuse energy-channel calibration
with system factor calibration (relevant for
standard-based quantification; for details see
section 5.12.3 and 5.12.4).
2
Set a calibration sample into the analysis
position.
Use a sample with any element in the mid
energy range (Mn – Zn) without peak overlap
and sufficient concentration. Preferably use
single element standards (e.g. pure copper) to
avoid peak misidentification, improve speed
and accuracy.
3
Set the beam current to produce an
intermediate count rate.
Use microscope controls. Set appropriate
magnification on a homogeneous sample
area.
4
Select element and spectral line for
calibration.
Use K
α
lines whenever possible.
5
Click
to check the current
energy channel calibration and deviation.
The test corresponds to the current
Pulse
throughput
and
Maximum energy
displayed
in the
EDS configurator
. The acceptable
deviation value is b 5 eV and - 5 eV.
If not, proceed with step 6.
6
Select
Pulse throughput
and
Energy
range
under
Settings to calibrate
.
7
Choose accuracy and click
.
Recommended value is
Medium.
8
Click
to save new calibration
data.
The calibration data is not user specific.
Changes in calibration will affect all users of
the ESPRIT software.
Содержание QUANTAX EDS
Страница 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Страница 6: ......
Страница 8: ...User Manual 2 Safety Information 8 ...
Страница 12: ......
Страница 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Страница 22: ......
Страница 24: ......
Страница 48: ......
Страница 50: ...User Manual 5 Step by Step Guides 50 ...
Страница 52: ...User Manual 5 Step by Step Guides 52 Fig 5 1 1 Preparatory steps for measurements with the ESPRIT software ...
Страница 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Страница 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Страница 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Страница 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Страница 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Страница 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Страница 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Страница 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Страница 85: ...5 12 Quantitative EDS Analysis of Bulk Samples QUANTAX EDS 85 Fig 5 12 2 Automatic standardless quantification ...
Страница 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Страница 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Страница 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Страница 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Страница 116: ......