5.7 Spectrum Acquisition
QUANTAX EDS
63
5.7
Spectrum Acquisition
This section describes the workflow for spectrum acquisition within the Spectra workspace. It is
recommended to acquire a preview spectrum before proceeding with analysis in various
workspaces.
Step
Examples/hints
Choose appropriate measurement
conditions on the electron microscope.
Refer to the previous sections 5.1, 5.2, and
5.3.
1
Select the
workspace.
2
Click the
button to acquire
a live spectrum.
The live spectrum appears in the spectrum
window. Click the
button
again to stop acquisition.
3
Click
on the
button to
set timing parameters in the
Preview
timing
submenu.
4
Click
on the
button
a) Set
Acquisition parameters
Set acquisition time.
Automatic
Fast
(50,000 counts) for major elements,
Precise
(250,000 counts) for minor elements,
Exhaustive
(1,000,000 counts) for elements
close to the detection limit.
Manual
If
Manual
is selected, the acquisition has to
be stopped manually by clicking the
button.
Real Time
The acquisition will stop after the time
entered in the dialog box has elapsed.
Live Time
The acquisition will stop after the dead
time corrected acquisition time has elapsed.
Counts
The acquisition will stop after the predefined
number of counts is recorded.
b) Set
Automatic quantification
If
Continuous/After acquisition
is selected,
the spectrum is quantified during/after
acquisition. Click to access the method
editor (see description in 5.12.1). Load a
quantification method using the
button.
c) Activate
Cyclic acquisition
This option acquires several spectra with
identical settings from the same area. Set
Cycle count and Pause [s]. Use
Add to
project
to automatically add acquired spectra
to project.
d)
Spectrum numbering
.
Automatic numbering
can be activated. The
numbering will start with the entered
Spectrum number
.
Содержание QUANTAX EDS
Страница 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Страница 6: ......
Страница 8: ...User Manual 2 Safety Information 8 ...
Страница 12: ......
Страница 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Страница 22: ......
Страница 24: ......
Страница 48: ......
Страница 50: ...User Manual 5 Step by Step Guides 50 ...
Страница 52: ...User Manual 5 Step by Step Guides 52 Fig 5 1 1 Preparatory steps for measurements with the ESPRIT software ...
Страница 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Страница 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Страница 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Страница 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Страница 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Страница 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Страница 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Страница 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Страница 85: ...5 12 Quantitative EDS Analysis of Bulk Samples QUANTAX EDS 85 Fig 5 12 2 Automatic standardless quantification ...
Страница 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Страница 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Страница 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Страница 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Страница 116: ......