User Manual
5 Step-by-Step Guides
106
5.19
Map Processing
This section describes the online and offline processing of HyperMap data (datacube/.bcf files*).
Step
Example/hints
1
Select the
workspace.
2
Use the
icon to open the periodic
table and identify elements.
Select elements by clicking on the
Auto ID
or
use the
Finder
in the Spectrum tab. Identified
elements are displayed in the map thumbnail
bar.
3
Select/Deselect element(s) in the
thumbnail bar by ticking/unticking the
boxes below the individual element
images to display their distribution in the
large/ mixed map image.
Optionally use
or
on the right of
the thumbnail bar.
4
Click to open the
Map display
settings
submenu and adjust
Alternatively, click with the left mouse button
into the Map display bar.
a)
Image filter
Applies filters on the overlay EM image.
b)
Map filter
Applies filters on the element distribution
map.
c)
Result types
Use
Online deconvolution
to display
deconvolved elements distribution maps.
d)
Color control
Use sliders to apply image brightness,
gamma correction and color saturation.
e)
Map color mixing.
Use
Enhanced
to improve the display of EM
image and element map overlay.
5
Click on
to switch to pseudo colors
display.
Choose one element only for pseudo color
display. Right click on the icon to adjust
Palette settings
. Note: If the map is not
quantified, the values are normalized to the
highest displayed intensity value.
6
Extract point/area spectra from the map:
i.
Right click on
to access
object tool list
ii.
Select the preferred tool
iii.
Click on
to activate the
object tool
iv.
Move the cursor to the area of
interest, press left mouse key and
drag cursor to expand object.
View and quantify the extracted point/area
spectra in the
Spectrum tab
.
Select point, line, rectangle, ellipse or
polygon.
Object spectra appear in
Spectrum tab.
Line profile appears in the
Line scan tab.
Enhance line profile statistics: use + and –
keys to set line width. Line scan can be
transferred to the Line Scan workspace via
the Project. To quantify a map or a line scan,
see section 5.18 or 5.16.
7
Click on
to generate a
MaximumPixelSpectrum
Identification of locally enriched elements
with low total concentration. The
MaximumPixelSpectrum is displayed in the
Spectrum
tab.
Содержание QUANTAX EDS
Страница 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Страница 6: ......
Страница 8: ...User Manual 2 Safety Information 8 ...
Страница 12: ......
Страница 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Страница 22: ......
Страница 24: ......
Страница 48: ......
Страница 50: ...User Manual 5 Step by Step Guides 50 ...
Страница 52: ...User Manual 5 Step by Step Guides 52 Fig 5 1 1 Preparatory steps for measurements with the ESPRIT software ...
Страница 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Страница 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Страница 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Страница 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Страница 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Страница 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Страница 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Страница 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Страница 85: ...5 12 Quantitative EDS Analysis of Bulk Samples QUANTAX EDS 85 Fig 5 12 2 Automatic standardless quantification ...
Страница 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Страница 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Страница 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Страница 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Страница 116: ......