User Manual
5 Step-by-Step Guides
100
5.16
Quantitative Line Scan (QLine)
This section describes how to quantify a line scan.
Step
Example/hints
1
Select the
workspace.
2
a
line scan or
Load
a stored line scan file (.rtl) from the
workspace
menu, or
retrieve it from the
Project
folder.
See section 5.15.
3
Click on the
button and
a)
Load…
an optimized quantification
method
Use the Scan spectrum of the Line scan to
optimize a quantification method. Save this
method to disk or add to project and drag and
drop from the project into the Profiles tab.
See section 5.12.1 for details. For electron
transparent samples use a suitable Cliff-
Lorimer or Zeta method.
b) Set
Quant range.
To enhance spectra statistics, use values 3, 5,
7, 9, 11, 13 to bin neighboring point spectra.
4
Click
to start the
quantification of the line scan database.
5
Use the
icon in the
Profiles
tab to
change profile display options.
Select from:
Counts
,
Net sum
,
Mass
percent
,
Mass percent (norm.)
and
Atomic
percent (norm.)
6
Use the workspace
icon to
When adding the data into the project, or
saving the line scan .rtl file format, the point
spectra for each point in the line scan can be
added to the
Project
by confirming the pop-
up window
Do you want to save point
spectra too?
a)
Save
the line scan
Use .rtl file format to save quantified line scan
data including scan image and point spectra.
b)
Add
data
to project
Image(s), composite element profiles and
scan image will be transferred.
Confirm the pop-up window
Do you want to
save point spectra too?
by clicking
to transfer spectral data.
c)
Add
data
to report.
7
Use the
Profiles
tab
icon to
a)
Save
data
Save as .bmp, .jpg, .png, .tif file format
including
Composite element profile
,
Scan
image
or
both
.
b)
Add
data
to project
Alternatively, right click into the profile and
use the local
Line scan
menu. Confirm the
pop-up window
Do you want to save point
spectra too?
by clicking
to transfer
spectral data.
c)
Add
data
to report.
Содержание QUANTAX EDS
Страница 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Страница 6: ......
Страница 8: ...User Manual 2 Safety Information 8 ...
Страница 12: ......
Страница 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Страница 22: ......
Страница 24: ......
Страница 48: ......
Страница 50: ...User Manual 5 Step by Step Guides 50 ...
Страница 52: ...User Manual 5 Step by Step Guides 52 Fig 5 1 1 Preparatory steps for measurements with the ESPRIT software ...
Страница 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Страница 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Страница 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Страница 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Страница 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Страница 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Страница 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Страница 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Страница 85: ...5 12 Quantitative EDS Analysis of Bulk Samples QUANTAX EDS 85 Fig 5 12 2 Automatic standardless quantification ...
Страница 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Страница 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Страница 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Страница 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Страница 116: ......