User Manual
5 Step-by-Step Guides
76
5.11.2
Maintain Standards Library for Electron Transparent Samples –
Cliff-Lorimer Factors
This guide describes how to check and calculate theoretical Cliff-Lorimer factors for standardless
quantification of electron transparent samples using the standards library dialog.
Step
Example/hints
1
Click the
icon in the bottom left
corner of the
Standards configurator
to
open the
ADMINISTRATION OF
STANDARD LIBRARY
dialog.
2
Check the detector geometry parameters
.
These parameters have to match the
parameters of measured data and/or settings
of your EDS configuration on the microscope.
A spectrum measured with the correct
system settings will also contain the
respective values. These can be checked in
the file or by double click on the text line
below the spectrum chart. The geometry
parameters of a measurement system can be
checked in the
workspace by
clicking on
and should only be
changed by Bruker service. If the geometry
parameters do not match, proceed with step
3. If they match, proceed with quantification
as described in section 5.13.1 or 5.13.2.
3
Use the
standards library
icon to
a)
Load
a standards library or
Load a previously saved *.esl file which
corresponds to detector geometry
parameters. Proceed with step 5.
b) Select
New
to create an empty
standards library
The question
“Do you want to load a
spectrum measured with corresponding
system settings?”
pops up.
i.
Select
Yes
If you are using the software as Data station
(not connected to a microscope). The
LOAD
SPECTRUM
dialog pops up. Load a spectrum
*.spx which was acquired with corresponding
detector geometry.
ii.
Select
No.
If you are using the ESPRIT software
connected to the microscope.
4
Check
NEW STANDARD LIBRARY
geometry parameters, enter a name and
click
.
Check or update values of Energy, Elevation
angle, Azimuth angle, Tilt angle, Beam
entrance angle. They should match the values
of your microscope/detector configuration
(see step 2).
5*
Click on
to review.
The
EDIT CLIFF-LORIMER FACTORS
dialog
pops up: Cliff-Lorimer factors for K-, L- and M-
lines are listed. Zoom the graph: left click into
it and use the arrow keys on keyboard.
Содержание QUANTAX EDS
Страница 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Страница 6: ......
Страница 8: ...User Manual 2 Safety Information 8 ...
Страница 12: ......
Страница 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Страница 22: ......
Страница 24: ......
Страница 48: ......
Страница 50: ...User Manual 5 Step by Step Guides 50 ...
Страница 52: ...User Manual 5 Step by Step Guides 52 Fig 5 1 1 Preparatory steps for measurements with the ESPRIT software ...
Страница 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Страница 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Страница 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Страница 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Страница 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Страница 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Страница 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Страница 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Страница 85: ...5 12 Quantitative EDS Analysis of Bulk Samples QUANTAX EDS 85 Fig 5 12 2 Automatic standardless quantification ...
Страница 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Страница 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Страница 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Страница 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Страница 116: ......