minispec.exe Software for Diagnostic and Service Purposes
E1400912_1_001
61
The command
Continue
is used to leave an intermediate process in order to continue
running the minispec application (e.g. during a waiting time, a data acquisition or a fit
display).
Shortcuts:
Standard Tool Bar:
Measurement Tool Bar:
Keys: ESC
The Calibrate Command
A typical minispec application detects NMR signals of the sample and then calculates the
result from the NMR signal. In some cases, the result can be directly calculated from the
measured signal (e.g. a relaxation time). In other cases the minispec applications may use a
reference method and calibration line to determine a sample property (e.g. oil, water or fat
content) from the NMR signal. As a consequence, the NMR signal must first be calibrated
with respect to sample properties. Select
Calibrate Sample
from the
Measurement
menu.
Shortcuts: Standard Tool Bar:
Measurement Tool Bar:
Keys: ALT+C
Running a minispec sample calibration leads the user through several steps. In most cases
some application specific values (see: The Application Parameter Table Command) need to
be defined.
To calibrate a minispec application, proceed as follows:
1. Prepare at least 3 samples of known content for measurement.
2. Perform test measurements with each calibration sample to adjust the acquisition
parameters from the
Acquisition Parameter Table
in the
Parameter
menu. Adjust the
receiver gain to the sample with the strongest signal.
3. Save the parameters and leave the parameter table.
4. Run the sample calibration and follow the prompts.
A flag in the
Info
Toolbar indicates a calibrated minispec application.
An uncalibrated minispec application is labeled with the following flag:
Содержание minispec mq20
Страница 1: ...minispec mq20 Toothpaste Analyzer User Manual Version 001 Innovation with Integrity AIC ...
Страница 5: ...Contents E1400912_1_001 5 Index 83 ...
Страница 6: ...Contents 6 E1400912_1_001 ...
Страница 14: ...Safety 14 E1400912_1_001 ...
Страница 18: ...Transport Packaging and Storage 18 E1400912_1_001 ...
Страница 22: ...Technical Data 22 E1400912_1_001 ...
Страница 72: ...minispec exe Software for Diagnostic and Service Purposes 72 E1400912_1_001 ...
Страница 76: ...Maintenance 76 E1400912_1_001 ...
Страница 78: ...Dismantling and Disposal 78 E1400912_1_001 ...
Страница 80: ...Contact 80 E1400912_1_001 ...
Страница 82: ...List of Figures 82 E1400912_1_001 ...
Страница 85: ...E1400912_1_001 85 ...
Страница 86: ...Bruker Corporation info bruker com www bruker com Order No E1400912 ...