Running the minispec mq20 Toothpaste Analyzer
E1400912_1_001
41
Afterwards switch back to the main menu of minispec Plus to start running measurements of
unknown samples.
7.4
Measurements of Unknown Samples
When the calibration(s) is(are) finished, one may switch to the “Measurement” section. Here
unknown samples are analyzed and measured according to the calibration curve selected by
the user.
• Enter the batch name of the sample to be analyzed.
• Select the corresponding calibration method. Note: once batch and calibration are chosen,
those are connected to each other. The software then remembers this connection when
doing further measurements.
• Depending upon the method, as it has been defined during the calibration, the sample
may need to be weighed. Enter the corresponding sample weight into the software or get
the mass transferred from the balance connected.
Содержание minispec mq20
Страница 1: ...minispec mq20 Toothpaste Analyzer User Manual Version 001 Innovation with Integrity AIC ...
Страница 5: ...Contents E1400912_1_001 5 Index 83 ...
Страница 6: ...Contents 6 E1400912_1_001 ...
Страница 14: ...Safety 14 E1400912_1_001 ...
Страница 18: ...Transport Packaging and Storage 18 E1400912_1_001 ...
Страница 22: ...Technical Data 22 E1400912_1_001 ...
Страница 72: ...minispec exe Software for Diagnostic and Service Purposes 72 E1400912_1_001 ...
Страница 76: ...Maintenance 76 E1400912_1_001 ...
Страница 78: ...Dismantling and Disposal 78 E1400912_1_001 ...
Страница 80: ...Contact 80 E1400912_1_001 ...
Страница 82: ...List of Figures 82 E1400912_1_001 ...
Страница 85: ...E1400912_1_001 85 ...
Страница 86: ...Bruker Corporation info bruker com www bruker com Order No E1400912 ...