
minispec.exe Software for Diagnostic and Service Purposes
E1400912_1_001
49
8.1.2.2
Changing Analyzer Type and Calibration Permission
Use the
Customize
button in the
configuration level
(see Selecting Applications and
Connections) to open a dialog box. In this window the analyzer type and the calibration
permission can be selected by checking the corresponding select box with a mouse click.
Two options, RELAXATION TIME and the NMR analyzer, require a license number to
operate. The number will be requested while connecting to the minispec (after you click the
OK
button in the Welcome Box) when one of these options is selected. If ordered with the
system, this license number is issued on the first page of this booklet, labeled as the ExpSpel
license number.
8.1.3
General Measurement Handling
The most important commands that are used to control measurements are: the
Run
command to start a measurement and the
Stop
command to interrupt a measurement in
progress. Measurements can be started through a number of commands and by activating a
number of software icons or features. For example, the general Run command can be started
through the
RUN
button from the top icon bar or bottom button panel, from the pull-down
menu
minispec
, or by simply pressing <
Enter
>. If the user has been given permission to
perform sample calibrations, then the
Calibrate
command is available. Other forms of
measurement are available for convenience, including a repetitive run command and a
command to detect the raw NMR signal.
The following sequence of actions is generally followed in preparation for measurements:
1. Turn on your minispec and your PC (if they were turned off).
´
Wait until the console CPU/Ethernet process boots, as indicated by the lighting of the
left (green) LEDs located on the front edge of the magnet and console.
2. Connect the PC to the minispec you want to use and configure the software,
e.g.
choose
applications or an analyzer type.
´
Wait until the right LED on the magnet unit is green. This can take up to 3 hours to
ensure optimum temperature stability. If the system was turned off for several days, or
it was moved to another place, insert the calibration sample with 0% solid content and
run the
Update Settings
routine.
3. Depending on the application you use (see your application description), start your
measurement or calibration with the
Run
or
Calibrate
command.
The software includes full automatic tuning routines to adjust the measurement settings, for
example pulse length, receiver gain, and magnetic field. You can tune these settings
periodically to ensure the instrument validity over time.
The
Daily Check
command should be executed once a day to check the most important
instrument settings and hardware functionality.
Every 24 hours, a yellow field on the right of the display will automatically indicate the need
for another
Daily Check
. Running the
Daily Ch
eck periodically ensures conformance to
Good Laboratory Practice (GLP).
Содержание minispec mq20
Страница 1: ...minispec mq20 Toothpaste Analyzer User Manual Version 001 Innovation with Integrity AIC ...
Страница 5: ...Contents E1400912_1_001 5 Index 83 ...
Страница 6: ...Contents 6 E1400912_1_001 ...
Страница 14: ...Safety 14 E1400912_1_001 ...
Страница 18: ...Transport Packaging and Storage 18 E1400912_1_001 ...
Страница 22: ...Technical Data 22 E1400912_1_001 ...
Страница 72: ...minispec exe Software for Diagnostic and Service Purposes 72 E1400912_1_001 ...
Страница 76: ...Maintenance 76 E1400912_1_001 ...
Страница 78: ...Dismantling and Disposal 78 E1400912_1_001 ...
Страница 80: ...Contact 80 E1400912_1_001 ...
Страница 82: ...List of Figures 82 E1400912_1_001 ...
Страница 85: ...E1400912_1_001 85 ...
Страница 86: ...Bruker Corporation info bruker com www bruker com Order No E1400912 ...