Delta Class SLA6900 UHP & SLA6800 Series Metal Sealed MFC's/MFM's
1-4
Section 1 Introduction
Installation and Operation Manual
X-TMF-SLA6900-SLA6800-MFC-eng
Part Number: 541B100AAG
August, 2009
Power Requirements
Watts, typ.
Watts, max.
Analog I/O option, no valve:
1.80
Analog I/O option, with valve:
2.60
4.00
DeviceNet I/O option, no valve:
3.80
DeviceNet I/O option, with valve:
4.90
7.60
Command-Setpoint Input (Analog I/O option only)
0-5 Vdc: Input will accept signals to 5.5 Vdc (110% F.S.).
Input resistance: 360 k ohm min.
Flow Output (Analog I/O option only)
0-5 Vdc into 2K ohm minimum load.
Output will indicate process variable up to 6 Vdc
(120% F.S.).
Valve Override Signal (Analog I/O option only)
Left floating/unconnected – instrument controls valve to command setpoint
Connected to signal at or above 5.0 Volts
– valve is forced open
Connected to signal at or below 0.0 Volts
– valve is forced closed
5 Volt Reference Signal (Analog I/O option & 15 PIN only)
For use with potentiometer command setpoint input
±0.2%, into 1 K ohm (minimum load)
Calibration-Select Signal (15 Pin Only)
Select one of ten gases and P.I.D. tuning settings in analog mode.