56 Cleaning the Vivo 40 and Replacement of Accessories
Vivo 40 users manual
Doc. 003821 En-Uk E-1
HA 20 Humidifier
The HA 20 humidifier should be cleaned, maintained and replaced in
accordance with the care provider's instructions.
3
At regular cleaning after normal
use, keep the rear lid or HA 20 heat-
ing unit attached. If attached, allow
the HA 20 humidifier to cool down
for ten minutes. Detach only the
HA 20 water chamber.
In case a more thorough cleaning is
needed, detach the rear lid or the
HA 20 heating unit.
4
Disconnect all electric cables.
5
Clean the outside of the Vivo 40
using a lint-free cloth and a mild
soap solution. If the surface of the
Vivo 40 needs to be disinfected, this
can be done with Virkon
®
or Gigas-
ept
®
.
6
Reconnect the patient cir-
cuit. Make sure all parts are dry
before the Vivo 40 is put into
operation.
For more information, see Breas HA 20 Humidifier User Manual.