40 Preparing the Vivo 40 for Use
Vivo 40 users manual
Doc. 003821 En-Uk E-1
6.2 Placing the Vivo 40
6.3 Connecting the Vivo 40 to the Mains Supply
To connect the Vivo 40 to the Mains Supply:
Read the chapter “Environmental Conditions” on page 10 carefully to
make sure all conditions are fulfilled and considered.
1
Place the Vivo 40 on a solid, flat
surface facing towards the patient.
The Vivo 40 should be placed lower
than the patient in order to prevent
the device from falling on the
patient, as well as preventing conden-
sated water to reach the patient.
2
Make sure that nothing can block
the patient air inlet at the rear of the
Vivo 40.
Read the chapter “Electrical Safety” on page 9 carefully to make sure all
conditions are fulfilled and considered.
1
Plug the power cord into the
power inlet of the Vivo 40.
2
Create a small loop on the
cable in order to prevent
stretching. Secure the power
cord using the cable holder.