61
Technical Specifications
Vivo 40 users manual
Doc. 003821 En-Uk E-1
10 Technical Specifications
10.1 Data
The Vivo 40 and it’s packaging do not contain any natural rubber latex.
S
ETTING
/V
ALUE
R
ANGE
/P
ERFORMANCE
R
ESOLUTION
Ventilation
modes
• PSV (Pressure Support
Ventilation)
• PCV (Pressure Control
Ventilation)
• CPAP (Continuous Posi-
tive Airway Pressure)
Patient modes
• Adult
• Paediatric
Target Volume
• Active
• Inactive
Device modes
• Clinical
• Home
EPAP 2 cmH
2
O to 20 cmH
2
O or
IPAP/IPAPmin -2 cmH
2
O
Tolerance: ±2% of the maxi-
mum value and ±10% of the
set value.
0.5 below 10 cmH
2
O
1.0 above 10 cmH
2
O
IPAP
Adult mode: 4 to 40 cmH
2
O
Paediatric mode: 4 to
25 cmH
2
O
Tolerance: ±2% of the maxi-
mum value and ±10% of the
set value.
0.5 below 10 cmH
2
O
1.0 above 10 cmH
2
O