58 Cleaning the Vivo 40 and Replacement of Accessories
Vivo 40 users manual
Doc. 003821 En-Uk E-1
8.2 Cleaning and Replacing the Patient Air Filters
Washable Filter (grey)
Replace the washable filter at least once a year. Wash the filter at least once a
week.
1
Wash the filter using warm water and a mild soap.
2
Rinse thoroughly.
3
Dry the filter by squeezing it out in a towel. Do not wring the filter.
Disposable Filter (white)
Replace the filter at least every 4th week, or more often when used in high pol-
lution or pollen-rich environments.
The patient air filters are located in
the filter cassette at the rear of the
ventilator. There are two types of fil-
ters: washable filter and disposable fil-
ter.
Do not wash or reuse the disposable filter.