Technical Specifications
Vivo 40 users manual
Doc. 003821 En-Uk E-1
Expiratory trig-
ger effort level
1 to 9, where 1 is the lowest
effort and 9 is the highest
effort setting.
1
Ramp function
On, Off, Disabled
HA 20 humidi-
fier
Settings: Off, 1 to 9, corre-
sponds to 10 to 30 mgH
2
O/l,
<100% RH. Heat-up time
from 23°C (73°F): less than
1 hour. Max gas temperature
at patient port: 43°C (109°F).
1
Audible alarm
level
1 to 9, where 1 is the lowest
and 9 is the highest volume
setting.
1
Maximum flow
> 200 litre/min
Maximum lim-
ited pressure
during single
fault condition
PCV, PSV:
60 cmH
2
O (Adult),
35 cmH
2
O (Paediatric).
CPAP: 30 cmH
2
O
Maximum steady
limiting pressure
Set IPAP + 10 cmH
2
O, toler-
ance: ±10%
Max flow in
CPAP mode
1/3 of max press.: 110 litre/
min
2/3 of max press.: 150 litre/
min
Max pressure: 155 litre/min
Breathing resist-
ance under sin-
gle-fault
4 cmH
2
O at 30 litre/min
6 cmH
2
O at 60 litre/min
Sound level at
10 cmH
2
O
Less than 30 dB(A)
Measured at 1 m
S
ETTING
/V
ALUE
R
ANGE
/P
ERFORMANCE
R
ESOLUTION