![Brainlab CRANIAL/ENT OPTICAL TRACKING Скачать руководство пользователя страница 78](http://html.mh-extra.com/html/brainlab/cranial-ent-optical-tracking/cranial-ent-optical-tracking_instrument-user-manual_2800942078.webp)
Step
1.
Fit the clamp
①
over the instrument.
NOTE: For the highest tracking accuracy, attach the clamp as close to the instrument tip
as possible, without touching the patient.
2.
Attach the array so that the longest arm of the array is in line with the instrument axis.
3.
Tighten the fix knob
②
by hand, to secure the instrument adapter clamp.
NOTE: Use the Tool for Instrument Adapter, if needed.
Arrays and Clamps
78
Instrument User Guide Rev. 2.6 Cranial/ENT Optical Tracking