5.5.3
Removing the Skull Reference Package
Introduction
The following sections describe the different methods for
Skull Reference Package
removal.
Standard Procedure
①
Figure 33
Step
1.
Unscrew and remove the
Skull Reference Array
.
2.
Loosen the tube cap nut
①
.
3.
Remove the screw with the screwdriver.
4.
Remove the
Skull Reference Base
by pulling it upwards.
REFERENCES
Instrument User Guide Rev. 2.6 Cranial/ENT Optical Tracking
73