Battery Replacement
Figure 13
Step
1.
Unscrew the three screws with the Torx screwdriver (size TX6) provided in the storage
case.
2.
Remove the lower housing and exchange the battery with a new battery. Make sure to in-
sert the battery with the correct polarity as indicated on the label.
3.
Attach the housing again and make sure to tighten all the screws with moderate force.
Warning
Do not change the Softouch battery within patient surroundings or while in contact with a
patient.
During the battery exchange, make sure that no liquids ingress inside the pointer housing.
Lithium batteries must be compliant with IEC 60086-4.
NOTE: The only service available for Softouch is the replacement of the battery. Dispose of the
pointer and/or the battery separately, according to your organization’s rules and local legislation.
Softouch Pointer
36
Instrument User Guide Rev. 2.6 Cranial/ENT Optical Tracking
Содержание CRANIAL/ENT OPTICAL TRACKING
Страница 2: ......
Страница 20: ...Available Instruments and Classification 20 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 28: ...Instrument Calibration Matrix 28 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 52: ...Electromagnetic Compatibility 52 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 86: ...Instrument Adapter Clamp Softgrip Size L XL 86 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 94: ...Ultrasound Calibration 94 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 102: ...Attaching the VarioGuide to the Head Holder 102 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 110: ......
Страница 111: ......
Страница 112: ...brainlab com Art No 60921 34US 60921 34US ...