Step
8.
Apply slight pressure to the
Skull Reference Base
to insert the pins
②
into the patient's
skull.
9.
Turn the tube cap nut
clockwise
to press the pins into the bone, securing the
Skull Refer-
ence Base
.
How to Attach the Unsterile Array
①
②
③
④
Figure 31
Step
1.
Bring the unsterile
Skull Reference Array
down over the pin of the connector interface
③
and tighten the array cap nut
②
.
2.
Loosen the tube cap nut
①
and the adjustment nut
④
to set the position of the array so
that it is visible to the camera.
3.
Securely tighten both
①
and
④
before starting registration.
NOTE: Do not unscrew any of the reference screws after patient registration. Only the array cap
nut may be unscrewed, to exchange the reference array.
REFERENCES
Instrument User Guide Rev. 2.6 Cranial/ENT Optical Tracking
71
Содержание CRANIAL/ENT OPTICAL TRACKING
Страница 2: ......
Страница 20: ...Available Instruments and Classification 20 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 28: ...Instrument Calibration Matrix 28 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 52: ...Electromagnetic Compatibility 52 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 86: ...Instrument Adapter Clamp Softgrip Size L XL 86 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 94: ...Ultrasound Calibration 94 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 102: ...Attaching the VarioGuide to the Head Holder 102 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 110: ......
Страница 111: ......
Страница 112: ...brainlab com Art No 60921 34US 60921 34US ...