10 Technical specifications
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
95
Mode icon
Description
CCS
ARC
CONTROL
Form of HF
voltage
Max. power output
Peak
voltage
Default values
Effect
Power range
Effect
Max. Watt
Modes Bipolar Coagulation
Micro forceps
sinusoidal
constant
-
0.1 W - 20 W
150 Vp
---
10
Forceps
forced
sinusoidal
modulated
-
1 W - 100 W
550 Vp
---
70
LIGATION
sinusoidal
modulated
-
200 W
190 Vp
---
---
TissueSeal
PLUS
sinusoidal
modulated
-
200 W
190 Vp
---
---
Bipolar
scissors
sinusoidal
constant
-
1 W - 120 W
200 Vp
---
60
Laparoscopy
sinusoidal
constant
-
1 W - 120 W
150 Vp
---
50
Bipolar
resection
sinusoidal
constant
-
1 W - 350 W
190 Vp
---
200
The max. values are not necessarily created at rated load resistance.