12 EMC
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
135
12. EMC
12.1.
Guidelines and manufacturer’s declaration in
accordance with IEC 60601-1-2 Section
6.8.3.201
Emission of electromagnetic interference (IEC 60601-1-2, Table 201)
The ARC 400 is intended for operation in an electromagnetic environment as described below. The customer or
user of the ARC 400 should ensure that it is operated in such an environment.
Interference emission measurement
Conformity
Electromagnetic environment
guideline
HF emissions according to CISPR 11
Group 2
The ARC 400 must emit
electromagnetic energy in order to
perform its intended function. Nearby
electronic devices may be affected.
HF emissions according to CISPR 11
Class A
The ARC 400 is suitable for use in
facilities other than those suitable for
a residental environment or those
connected directly to the public
power grid, which also supplies
power to buildings used for
residential purposes.
Emission of harmonics according to
IEC 61000-3-2
Classes A and D
Emission of voltage fluctuations and
flickere according to IEC 61000-3-3
Conforms