![Bowa ARC 400 Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/bowa/arc-400/arc-400_operating-manual_2797070094.webp)
10 Technical specifications
94
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Mode icon
Description
CCS
ARC
CONTROL
Form of HF
voltage
Max. power output
Peak
voltage
Default values
Effect
Power range
Effect
Max. Watt
Modes Monopolar Coagulation
Argon flex.
pulse
pulsed
modulated
1
2
3
1 W – 80 W
4.4 kVp
2
20
Resection
sinusoidal
modulated
-
1 W -120 W
2.6 kVp
---
60
Cardiac
Mammary
sinusoidal
modulated
-
1 W - 60 W
2.3 kVp
---
15
Cardiac
Thorax
sinusoidal
modulated
-
1 W - 100 W
2.3 kVp
---
40
SimCoag
sinusoidal
modulated
pulsed
modulated
pulsed
modulated
1
2
3
1 W - 120 W
1.5 kVp
2.3 kVp
4.6 kVp
2
60
Gastro Coag
sinusoidal
modulated
1
2
3
1 W - 50 W
2.3 kVp
2.6 kVp
3.1 kVp
2
15
Laparoscopy
sinusoidal
modulated
-
1 W - 120 W
1.8 kVp
---
60
Modes Bipolar Cutting
Standard
Yes
Yes
sinusoidal
constant
-
1 W - 200 W
400 Vp
---
100
Bipolar
resection
Yes
Yes
sinusoidal
constant
1
2
3
250 W
500 Vp
2
---
Bipolar
scissors
sinusoidal
constant
-
1 W - 120 W
200 Vp
---
60
Modes Bipolar Coagulation
Standard
forceps
sinusoidal
constant
-
1 W - 120 W
150 Vp
---
40
Standard
forceps
AUTO
sinusoidal
constant
-
5 W - 120 W
150 Vp
---
40