10 Technical specifications
90
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Programs
Number of programs in the device
400
Default programs, factory set
Individually programmable
Program number and data shown on the display
Neutral electrode monitoring
EASY
(E
lectrode
A
pplication
Sy
stem)
Display indication of one-piece or split or Baby
electrode in the main and neutral electrode menu
Contact resistance between individual sections of
split neutral electrodes shown on display with
column for control
Lead resistance shown on the display when a
one-piece neutral electrode is used
Maximum allowable resistance between the
sections of a split electrode
300 Ohm
Warning signal for hazardous conditions beneath
the neutral electrode
Visual, acoustic
Warning messages on display:
Text messages with further
information
Safety features
ISSys
(I
ntegrated
S
afety
Sys
tem)
Continuous monitoring of HF leakage current with
fault indication
Text messages with further
information
Dosage monitoring with fault indication on the
display
Continuous self-test
Continuous status indication on the display
Operating errors shown on the display
Text messages with
further information
System faults shown on the display
Text messages with
further information
Documentation
Data acquisition and storage in the device
System information with
date
Fault states
Operating errors
Data retrieval via the display
Text messages with further
information