10 Technical specifications
110
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Monopolar Coagulation – Forced non cutting
Measurement of resistances
Diagram of power output P [W] as a
function of the load resistance R [
] for
the setting "Monopolar Coagulation
Forced non cutting"
= 40 W
Measurement of resistances
Diagram of power output P [W] as a
function of the load resistance R [
] for
the setting "Monopolar Coagulation
Forced non cutting"
= 80 W
Diagram of power output P [W] as a
function of the setting "Monopolar
Coagulation Forced non cutting"
Rated load resistance = 1000
HF output voltage U [Vp] for the setting
"Monopolar Coagulation Moderate"
(idle mode)
= 3500 Vp