![Bowa ARC 400 Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/bowa/arc-400/arc-400_operating-manual_2797070029.webp)
5 Operation
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
29
5. Operation
5.1.
Connecting power
The line voltage must match the voltage specified on the rating label.
Connect the power cord to the generator and plug the cord into a grounded AC
power outlet.
Using the power switch, switch the device on.
The device performs a functional test.
The bars of the activation display light up, while the connector surrounds light
up orange.
Full functionality of the loudspeaker is indicated by the start melody.
The individually configurable start screen appears if it has been set up.
5.2.
Program overview
5.2.1. Display
Figure 5-1: Main screen
The status bar is located at the top of the display.
The settings of the four socket connectors are shown below the status bar.
Settings can be configured for each of the connectors.
The "Effect" button is used to set the effect of electrosurgical cutting.
The "Pedal" button allows the activation of specific functions to be assigned to the
foot pedal.
The "Mode" button allows the desired type of current to be selected.
The "max. Watt" button is used to set the maximum output power.