5 Operation
68
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
The volume of the activation signal should be increased as necessary for
use in relativey noisy surrouondings. The alarm tones have a minimum
volume and limited changeability.
Mode
Category
Frequency (Hz) Signal type
Monopolar Cut
Activation tones
635
Continuous sound
Monopolar Coag
Activation tones
475
Continuous sound
Bipolar Cut
Activation tones
565
Continuous sound
Bipolar Coag
Activation tones
505
Continuous sound
Sim Coag
Activation tones
755
Continuous sound
GastroCut
/LIGATION end
Alarm tones
-
alternating sound
Error
Alarm tones
-
Signal tone
Warning
Alarm tones
-
Signal tone
Note
Alarm tones
-
Signal tone
5.10.4. "System messages" dialog
In the "System messages" dialog, it is possible to open the saved system
messages which have occurred since switching on the HF device.
These messages are not saved when switching off the HF device.
Opening saved system messages:
1. Select a system message.
2. The selected system message is displayed again with
"
?
"
.
3. Press
"
OK
"
to return to the overview.
Opening the instructions of use:
1. Press the
"
instructions of use
"
button.
The instructions of use are displayed chapter for chapter
2. Use the arrow buttons to select the desired chapter.
3. Press
"
OK
"
to open the selected chapter.
Use the arrow buttons to open the individual pages of the chapter.