- 53 -
4.4.11 Graphics
The „Graphics“ tab is used to define what is to be recorded and at what sampling rate.
„meassurand“
F[kN] = Force in kN or Lbf
S [mm] = Travel in mm or inches
I [A] = Current in amperes
U [V] = Voltage in volts
„sampling rate“
Defines the default sampling rate at which the curve is recorded; default
333 Hz.
„lag
max
“
Defines the time for which graphic recording continues after the target
criterion has been reached.
Recommended max. value 0.250 seconds, default: 0.150 seconds
„record time
max
“
Shows how many seconds can be recorded per setting for the selected
criteria.
If the recording time is not sufficient, the beginning of the recording is
discarded.