50
Voltage Monitor
This function is able to monitor the voltage output using Pin 11 and one
of the ground pins (i.e. Pin 22), which can be connected to a digital
voltage meter (DVM) or other voltage monitoring device, as shown
below. The output control must be in analog mode to use this function.
The monitoring of the output voltage range (which reflects 0 to full scale
of the power supply’s output voltage) can be selected between 0-5 V
and 0-10 V.
Figure 3.8 - Analog Voltage Monitor Diagram
Follow the steps below to configure the voltage program.
1.
Press the
button and press the
button one time until
CONFIG
is blinking and press
.
2.
Press
button three times until
EXTCTRL
is blinking and
press
.
3.
Press
button four times to select
MON RNG
and Press
.
4.
Now use the rotary knob to select
5V
or
10V
and press
.
5.
Press
several times to exit the menu setting.
Current Monitor
This function is able to monitor the voltage output using Pin 24 and one