95
Follow the steps below to perform the current calibration:
1.
Connect the DMM to the output of the power supply as shown
in the figure below:
Figure 5.2 – Current Calibration Diagram
2.
Set the DMM to a DC current measurement. Select the
CURRENT
parameter in the
CAL
menu and press
. The
following will be displayed:
The
P: 1/4
signifies there are 4 points to be calibrated, and the
current calibration point is 1.
3.
Input the current reading from the DMM in the
METER
parameter by using the numeric keypad and press
.