Fault Location Mode
59
Figure 22 Changed Max Distance Warning Message
Calibrate the Site Analyzer
If you need to use an extension cable or a phase stable
cable, connect that cable to the antenna test port on
the Site Analyzer and then connect the Bird Calibra-
tion Combination to the end of the extension cable.
Follow the procedure in Chapter 2, Calibration,
page 17.
Note:
After calibration, with the load still con-
nected, the output must be below –45 dB. If it is not,
there is a problem. It is common to see spikes from –60
to –45 dB. These can be ignored.
Remove the Cal Combo from the cable.