Bird Site Analyzer
18
Figure 2 Calibrating the Site Analyzer
8
800.00
MHz
1600.00
Calibration
Calibration: OFF
Load
Short
Open
10.00
1.00
V
S
W
R
(
0.00
/
)
6
7
1
5
2
3
4
Step
Action
1
Press the Calibrate key
2
Connect the Cal Combo “Open” connector to
the unit antenna port
3
Press the Open soft key then wait for the
unit to beep and display Done
4
Connect the Cal Combo “Short” connector to
the unit antenna port
5
Press the Short soft key then wait for the
unit to beep and display Done
6
Connect the Cal Combo “Load” connector to
the unit antenna port
7
Press the Load soft key then wait for the
unit to beep and display Done
8
Remove the Cal Combo from the unit